The probe tip input is equivalent to 100k resistor in parallel with about 2.3pF capacitance at low frequency.  
The probe tip input is equivalent to 100k resistor in parallel with about 2.3pF capacitance at low frequency.  
With the 10x or 100x attenuator tips, the probe input is equivalent to 1M paralleled with 2pF or 1.7pF, respectively.
With the 10x or 100x attenuator tips, the probe input is equivalent to 1M paralleled with 2pF or 1.7pF, respectively.
== Differences between S3 and S3A ==
The S-3 sampling head has been redesigned to improve baseline shift due to cable flexing. The improvement is roughly ten-to-one. All specifications on the S-3A remain the same as for the S-3. The major differences in design involves the gate board in the box part of the probe, but some changes are also made to the circuits in the body of the probe.